
|
|
Setup & PrincipleThe setup used for measuring the power actually consumed in radio-frequency excited, capacitively coupled plasmas is presented in the figure below: |
|
the L-C circuit represents the cell parasitics, | |
|
the Ls-Cs circuit is the shunt circuit connected externally and | |
|
the RL-CL circuit accounts for the resistive losses in the cell and the shunt circuit. |
Vn and In are the voltage and current waveforms measured using the voltage and current probe, while Ven, Ien are the equivalent voltage and current waveforms at the RF electrode.
The electrical analysis of the circuit shows that that the nth harmonic of the complex electrode voltage Ven and the complex electrode current Ien are related to the nth harmonic component of the measured complex voltage Vn and the measured complex current In by the relations (absence of shunt circuit):
|
Ven=Vn-jωnLIn Ien=(1-ωn2LC)In-jωnCVn |
For the actual cell parameters, the parasitic currents are large and the equivalent magnitudes are very sensitive even to small errors in the measured values. In fact, the electrode current Ien is a very small quantity compared the measured current In. This is because the actual electrode current results as the difference between two large numbers: the measured current minus the displacement current jωnCVn due to the capacitance between the RF electrode and the grounded shield. To avoid this difficulty the shunt circuit consisting of the capacitor Cs and the inductor Ls is connected to the electrical circuit. The shunt is tuned by exciting the empty cell with a sinusoidal voltage and adjusting the capacitor Cs until the measured current is nulled. In this ideal situation the net reactance jωLs+1/jωCs of the shunt cancels the net reactance of jωL+1/jωC of the cell equivalent circuit.
The shunt and the cell however are not pure reactances. They have also stray resistive components which cause additional losses upon them. Therefore the shunt does not exactly nulls the measured current. It is observed that the current takes a minimum value in the tuned situation. The remaining residual current is ~3.5% of the current without the shunt present. The attachment of the shunt circuit results to additional and considerable ohmic parasitics.
Measurements of this residual current have lead to the inclusion of this stray resistive component of the cell - shunt in the equivalent circuit. Namely, with the shunt properly tuned, the input impedance V1/I1 of the empty cell and shunt circuit is independent of the excitation voltage. Moreover, this residual current is in phase with voltage at the point of measurement when the cell is driven with a sinusoidal voltage. This behavior is as if the residual current flows, in parallel to the shunt circuit, from the RF power lead to the ground. This is taken into account by introducing a resistance RL in the equivalent circuit parallel to the shunt. The additional capacitor CL is introduced to account for the existence of the dc bias of the powered electrode which is due to the capacitive coupling of the power generator to the cell. This capacitance is considered to be infinite and is not involved in the calculations.
Finally, taking into account the shunt elements and the resistance RL the equations used for converting the measured values to the equivalent values at the electrode are given by the relations:
|
Ven=Vn-jωnLILn Ien=(1-ωn2LC)ILn-jωnCVn where ILn=In-Vn/RL-jωnCsVn/(1-ωn2LsCs) |
Experimental determination of the circuit elements
|
|
|
The determination of the exact values of the electrical components that are involved in the equivalent circuit combined to the measured values Vn, In can lead to the calculation of the voltage Ven and the current Ien at the surface of the RF electrode. The values of the cell capacitance C and cell inductance L are determined by exciting the empty cell with the shunt circuit disconnected at different driving frequencies. The results of the cell impedance that is calculated from the measured values of voltage and current, are further used to calculate the inductive L, capacitive C and resistive R components of the cell.
The experimental measurements and the fitting with the RLC model used for the calculation of voltage and current waveforms on the RF electrode are shown in the figure above. The excellent agreement between the experimental and the RLC model results, confirms the applicability of the equivalent circuit.
These results are applied to derive the following electrical characteristics for the chamber used in this example: inductance L=40nH, capacitance C=285pF, resistance R=0.8Ω and a resonance reactor frequency of 43MHz.
Furthermore, the resistance RL that has been introduced to the equivalent circuit to counterbalance the remaining residual current can be calculated by using the following procedure. The shunt circuit is connected to the circuit, is tuned and the cell is excited at different driving voltages. The values of voltage and current are recorded and plotted in the figure presented above. The slope of the curve result to the calculation of the resistance RL=1700 Ω.
In addition the values of the shunt circuit elements Cs and Ls are known as the capacitor is scaled and the inductor has fixed values. As the values of all the elements that are involved to the equivalent circuit can be determined, the values of the voltage Ven and current Ien at the surface of the RF electrode can be calculated. These values can be further used for the calculation of:
|
A complete description of the method as well as the extension to frequencies higher than 13.56 MHz are given in the following PTLUP publications:
"Power
dissipation and impedance measurements in radio frequency discharges"
N. Spiliopoulos, D. Mataras, D. E. Rapakoulias
J. Vac. Sci. Technol. A 14, 2757 (1996)
©PDF
"Frequency
variation under constant power conditions in hydrogen radio frequency
discharges"
E. Amanatides and D. Mataras
J. Appl. Phys. 89, 1556 (2001)
©PDF
Applications of the method are presented in a number of publications as well (see publications list).
Current and voltage waveforms - harmonics
The usefulness of the method is shown in the above figure.
The black curves in figures a, b represent the voltage and current waveforms as measured. The measurements are in an Argon discharge at a gas pressure of 100 mTorr, frequency of 13.56 MHz and excitation peak to peak voltage of 100 Volts.
The red curves represent the voltage and current waveforms transformed by applying the equivalent circuit for the RF electrode.
The measured and the calculated waveforms of the voltage differs only in phase, the amplitude being almost the same. In addition the contribution of higher harmonics is almost negligible being less than 2% of the fundamental frequency. On the other hand, there is a significant difference between the current waveform at the point of measurement and at the RF electrode. The current waveform at the RF electrode that represents the real discharge current differs in phase, amplitude and distribution of harmonics compared to the measured current. The difference in phase is the result of the delay that is introduced by the current probe. The difference in amplitude is the result of the current flow though the shunt circuit and the flow from the RF electrode to the ground. The strong hanarmonicity (about 20 % of the fundamental frequency, figure d) is the result of the non-linear response (non-linear impedance) of the discharge.